B l u e B o x – C a l i b r a t i o n p H S e n s o r
1.2 pH Sensor Calibration Reset to Default
Calibration Menu pH Sensor
Reset cal.
to default mV calibration
to default pH calibration
Switches back to the Calibration menu pH sensor.
<–
1.3 pH Sensor Offset
Calibration Menu pH Sensor
Slope
Input minus value, each 0,1 mV
Input plus value, each 0,1 mV
Switches back to the Calibration menu pH sensor without saving the entry.
Cancel
Saves the entry and switches back to the Calibration menu pH sensor.
Set
GO Systemelektronik GmbH
Reset calib ration
to default mV calibration
to default pH calibration
Default mV calibration, measurement value in mV
Offset = 0 | Slope) = 1
Switches back to the Calibration menu pH sensor.
7 pH ≙ 0 mV | Slope = 59,16 mV/pH unit
Default pH calibration, measurement value in pH
Switches back to the Calibration menu pH sensor.
0.3 mV
Cancel
input offset value = measured mV at pH 7
Faluner Weg 1
24109 Kiel
www.go-sys.de
<-
Offset
Set
Germany