Determining The Approximate Pneumatic Operating Time For The Oxylog 3000 - Dräger Oxylog 3000 Instructions For Use Manual

Emergency and transport ventilator
Hide thumbs Also See for Oxylog 3000:
Table of Contents

Advertisement

Determining the approximate pneumatic operating time for the
Oxylog 3000
Example for supply of medical gas:
– Cylinder pressure measured on the pressure
gauge of the pressure reducer: 2000 kPa
– Liquid capacity of the O
Supply of medical gas:
2.1 L x 2000 kPa = approximately 420 L
Example for pneumatic operation time:
– IPPV mode, frequency 10 breaths /min,
VT = 1 L, O
= 100%
2
– Minute volume = 10 breaths /min x 1 L =
10 L/min
Medical gas supply [L]
Operation time =
(MV +0.5*) [L/min]
* Calculated with average gas consumption of
ventilator: 0.5 L/min
420
Operation time =
10.5
The pneumatic operation time increases when
Oxylog 3000 operates with O
than 100 % O
as ambient air is drawn into the
2,
device.
The amount of gas from the high-pressure supply,
which is currently being consumed, is indicated by
the Oxylog 3000 in the lower left section of the
information window in L/min. This display is over-
written when a higher priority message is activated.
Instructions for Use Oxylog 3000 SW 1.n
cylinder: 2.1 L
2
= approx. 40 minutes
concentration of less
2
Example:
A
A O
consumption = 2.5 L/min
2
Getting Started
39

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents