1.6 Wiring

1.6.1 tM-TH8 Wiring

1.6.2 Wiring Recommendations

Use 26-12 AWG wire for signal connections.
Strip the wire to a length of 7±0.5mm.
Use a crimp terminal for wiring.
Avoid high-voltage cables and power equipment as
much as possible.
For RS-485 communication, use insulated and
twisted pair 24 AWG wire, e.g. Belden 9841.
12
tM-TH8 User Manual, Rev: A1.1 7MH-018-A11