Troubleshooting; System Error Messages - Endress+Hauser PROFIBUS Gammapilot M FMG60 Operating Instructions Manual

Radiometric measurement
Hide thumbs Also See for PROFIBUS Gammapilot M FMG60:
Table of Contents

Advertisement

Troubleshooting

78
7
Troubleshooting
7.1

System error messages

7.1.1
Error signal
Errors occuring during commissioning or operation are signalled in the following way:
• Error symbol, error code and error description on the display and operating module
• PROFIBUS PA signal:
– MAX: +99999
7.1.2
Last error
The last error is displayed in the "diagnostics" (*A) function group in the "previous error"
(*A1) function. This display can be deleted in the "clear last error" (*A2) function.
7.1.3
Types of error
Type of error
Symbol
Alarm (A)
continuous
Warning (W)
flashing
Alarm/Warning (E)
You can define whether the error should behave as an alarm or as a warning.
Gammapilot M FMG 60 - PROFIBUS PA
Meaning
The output signal assumes a definite value:
• MAX: +99999
Additionally, an error message is displayed.
The device continues measurement. An error message is displayed
(alternating with the measured value).

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hart gammapilot m fmg60

Table of Contents