Wiring In Terminal Compartment A - Endress+Hauser PROFIBUS Gammapilot M FMG60 Operating Instructions Manual

Radiometric measurement
Hide thumbs Also See for PROFIBUS Gammapilot M FMG60:
Table of Contents

Advertisement

Gammapilot M FMG 60 - PROFIBUS PA
4.8

Wiring in terminal compartment A

"
Caution!
Before connection please note the following:
• When using the instrument in hazardous
areas, make sure to comply with national
standards and the specifications in the Safety
Instructions (XA's). Make sure you use the
specified cable gland.
• The supply voltage must comply to the data
on the nameplate.
• Switch off power supply before connecting
the instrument.
• Connect equipotential bonding to the outer
tranmsitter ground terminal and to the ground
terminal of the water cooling jacket (if
present) before connecting up the instrument
(section "Potential equalisation").
• Connect protective earth to the protective
earth terminal (section "Terminal
assignment").
• According to IEC/EN 61010 a suitable power
switch has to be provided for the instrument.
• The cable isoloations must comply with the
supply voltage and the overvoltage category.
• The temperature resistance of the connecting
cable must comply with the ambient
temperature.
The procedure
1.
Using a 3 mm (7/64") Allen wrench, losen
the cover clamp for the lid.
2.
Unscrew the lid of the terminal
compartment.
3.
Push the power cable and (if required) the
signal cable) through the appropriate cable
glands or cable threads.
4.
Wire up according to the terminal
assignment diagram.
5.
Tighten the cable glands or threads.
6.
Screw the lid securely back onto the
termial compartment.
7.
Adjust the cover clamp so it is set over the
lid and tighten.
A
Á
À
Â
Ã
L1
N
+
-
L+
L-
Ä
Å
Æ
L00-FMG60xxx-04-00-00-xx-005
Wiring
25