Endress+Hauser PROFIBUS Gammapilot M FMG60 Operating Instructions Manual page 68

Radiometric measurement
Hide thumbs Also See for PROFIBUS Gammapilot M FMG60:
Table of Contents

Advertisement

Commissioning
ENDRESS + HAUSER
+
E
ENDRESS + HAUSER
+
E
ENDRESS + HAUSER
+
E
68
"Full calibration" (*16)
"Empty calibration" (*18)
full calibr.
*16
33 cps
empty calibr.
*18
2548 cps
The pulse rate of the respective calibration point is displayed in these functions. The displayed value
must be confirmed by pressing "E".
"-1" indicates, that no background calibration is present yet. In this case there are two options:
• either return to the "calibration" (*15) function and restart the calibration
• or enter a known or calculated pulse rate (manual calibration)
"Next point" (*19)
next point
*19
Âno
yes
This function is used to specify, if a further calibration point is to be entered or not.
Selection:
• no
This option must be selcted, after both calibration points have been entered. After this selection
the Gammapilot M returns to the group selection and the calibration is completed.
• yes
This option must be selected, if only one calibration point has been entered yet. After this
selection the Gammapilot M returns to the "calibr. point" (*13) function and the next point
can be entered.
Gammapilot M FMG 60 - PROFIBUS PA