Gammapilot M FMG 60 - PROFIBUS PA
ENDRESS + HAUSER
+
E
!
ENDRESS + HAUSER
+
E
"Reference pulse rate" (*1F)
ref. pulse rate
*1F
31826 cps
This function displays the reference pulse rate I
points. The value can not be edited.
Note!
I
is the pulse rate for the tube being empty (theoretical reference value). Generally, its value is
0
significantly higher than any real pulse rate occuring during the measurement.
"Next point" (*19)
next point
*19
Âno
yes
This function is used to specify, if a further calibration point is to be entered or not.
Selection:
• no
This option must be selected, if no furhter calibration point is to be entered or changed. After this
selection the Gammapilot M returns to the group selection and the calibration is completed.
• yes
This option must be selected, if a further calibration point is to be entered or changed. After this
selection the Gammapilot M returnt to the "calibr. point" (*1A) function and the next point
can be entered or changed.
6.5.4
Linearisation (for concentration measurements)
If the concentration is to be measured in a unit other than the "density unit" (*06), a linearisation
must be performed after the calibration. This can be done in the "Linearisation" (*4) function
group. The functions of this group and the linearisation procedure are described in the operating
manual BA 287F, "Gammapilot M - Description of Instrument Functions". This document can be
found on the supplied CD-ROM of the "ToF Tool - FieldTool Package".
6.5.5
Additional settings
After the calibration has been finished, the Gammapilot M outputs the measuring value through the
PROFIBUS PA interface.
Many further functions are available for optimizing of the measuring point. They can be
parametrized as required. A detailed description of all instrument functions is given in the operating
manual BA 287F "Gammapilot M - Description of Instrument Functions". This document can be
found on the supplied CD-ROM of the "ToF Tool - FieldTool Package".
, which results from the currently active calibration
0
Commissioning
77

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hart gammapilot m fmg60

Table of Contents