Wiring
26
4.9

Wiring in terminal compartment B

"
Caution!
Before connection please note the following:
• Connect equipotential bonding to tranmsitter
ground terminal and to the ground terminal of
the water cooling jacket (if present) before
connecting up the instrument (see above,
section "Potential equalisation").
• The cable isoloations must comply with the
supply voltage and the overvoltage category.
• The temperature resistance of the connecting
cable must comply with the ambient
temperature.
The procedure
1.
Unscrew the lid of the terminal
compartment.
2.
Push the following cables through the
appropriate cable glands or threads:
– signal cable (if the signal output is located
in terminal compartment B)
– PT-100 cable (if present)
– cascading cables (input and/or output, if
required)
3.
Wire up according to the terminal
assignment diagram.
4.
Tighten the cable glands or threads.
5.
Screw the lid securely back onto the
terminal compartment.
Gammapilot M FMG 60 - PROFIBUS PA
B
À
Á
OFF
ON
SIM WP
PT100
Â
13
9 10 11 12
IN CASCADE OUT
+ -
+ -
-
+
3 4
5 6
7 8
0
Ã
Ä
L00-FMG60xxx-04-00-00-xx-006