6.3

Basic setup

6.3.1
"Present date" (*01)
present date
*01
17.11.04 10:30
dd.mm.yy hh:mm
Date and time of the basic setup are specified in this function.
Format:
dd.mm.yy hh:mm
where:
– dd: the day ("01" ... "31")
– mm: the month ("01 ... "12")
– yy: the year (e.g. "04" for 2004)
– hh: the hour ("00" ... "23")
– mm: the minute ("00" ... "59")
Each of these values must be confirmed by "E" after it has been entered.
6.3.2
"Isotope" (*03)
isotope
*03
Â137Cs
60Co
no compens.
This function is used to specify which isotope is used for the measurement.
The Gammapilot M needs this information for the decay compensation.
Selection:
137
Cs
60
Co
• no compensation