Gammapilot M FMG 60 - PROFIBUS PA
ENDRESS + HAUSER
–
+
E
6.4.3
Full and empty calibration or covered and free calibration
Excerpt from the operating menu
The following excerpt from the operating menu shows how the full and empty calibration (for level
measurements) or the covered and free calibration (for level limit detection) are entered.
The individual functions are explained in the sections below.
The functions are only accessible after the background calibration has been performed.
*14
value full
[%]
target value at
calibr. point
*13
calibr. point
0: full/covered
1: empty/free
*17
value empty
[%]
target value at
calibr. point
Note: The functions "value full" (*14) and "value empty" (*17) are only present if the option "level" has been selected in
the "measuring mode" (*05) function.
"Calibration point" (*13)
calibr. point
*13
Âfull/covered
empty/free
This function is used to select which calibration point ("full/covered" or "empty/free") will be
entered.
Selection:
• full/covered
• empty/free
E
*15
*11
calibration
avg. pulse
1
0: stop/edit
rate
1: start
0
E
*15
*11
calibration
avg. pulse
1
0: stop/edit
rate
1: start
0
Commissioning
*16
full calibr.
[c/s]
-1: not calibrated
*19
next
point
Group
0: no
selection
1: yes
*18
empty calibr.
[c/s]
-1: not calibrated
L00-FMG60xxx-05-00-00-en-014
65