Diagnosis And Troubleshooting; General Troubleshooting - Endress+Hauser Proline Promag W 800 Operating Instructions Manual

Electromagnetic flowmeter
Hide thumbs Also See for Proline Promag W 800:
Table of Contents

Advertisement

Proline Promag W 800
For output signals
Error
Signal output outside the valid range
Device shows correct value on local display, but
signal output is incorrect, though in the valid
range.
Device measures incorrectly.
Measuring device not in smartphone or tablet
live list
Device not responding via SmartBlue app
Login via SmartBlue app not possible
Device cannot be operated via SmartBlue app
No write access to parameters
For access
Error
Write access to parameter not possible.
Write access to parameter not possible.
Endress+Hauser
12

Diagnosis and troubleshooting

12.1

General troubleshooting

For local display
Fault
Local display remains dark for
longer than 5 seconds when
touched.
Possible causes
Main electronics module is defective.
Parameter configuration error
Configuration error or device is operated outside
the application.
Bluetooth communication set to "on touch"
No Bluetooth connection
Device is being put into operation for the first
time
Incorrect password entered
Password forgotten
Hardware write protection enabled
Possible causes
Hardware write protection is enabled.
Current user role has limited access
authorization.
Possible causes
Supply voltage does not match the
voltage specified on the nameplate.
Supply voltage has incorrect
polarity.
The connecting cables are not
connected correctly.
No battery pack inserted or
connected.
No buffer capacitor inserted or
connected.
Device is not powered from the
mains.
Order spare part →  92.
Check and adjust parameter configuration.
1. Check and correct parameter configuration.
2. Observe limit values specified in the
"Technical Data".
1. Check whether the Bluetooth logo is visible on
the local display or not.
2. Touch the display for 5 seconds so that a
measured value is displayed.
Enable Bluetooth function on smartphone or
tablet.
The device is already connected to another
smartphone/tablet.
Enter initial password (device serial number)
and change.
Enter correct password.
Contact Endress+Hauser Service.
• Check user role
• Enter the correct customer-specific access
code
• Hardware write protection via DIP switch
Set the write protection switch on the back of
the display to the ON position →  11.
1. Check user role →  11.
2. Enter correct customer-specific access code
→  11.
Diagnosis and troubleshooting
Remedial action
Apply the correct supply voltage
→  53.
Reverse polarity of supply voltage.
Check the cable connection and
correct if necessary.
Insert or connect battery pack.
Insert or connect buffer capacitor.
Touch the display for 5 seconds
→  72.
Remedial action
Remedial action
75