Gammapilot M FMG 60 - PROFIBUS PA
!
"
!
5.4

Lock/unlock configuration

5.4.1
Software security locking
Enter a number ≠ 2457 into the "unlock parameter" (*A4) function in the "diagnostics" (*A)
function group.
The
symbol appears on the display. Inputs are no longer possible.
If you try to change a parameter, the device jumps to the "unlock parameter" (*A4) function.
Enter "2457".
Now change the parameters.
5.4.2
Hardware security locking
Press S, O and F simultaneously.
Inputs are no longer possible.
If you try to change a parameter, the following appears:
Press S, O and F simultaneously. The "unlock parameter" (*A4) function appears.
Enter "2457".
Now change the parameters.
Note!
A hardware locking can only be unlocked again via the display by pressing the O, S and F keys at
the same time again. It is not possible to unlock the hardware by communication.
5.5

Reset to the default configuration

It is advisable to reset the customer parameters if you want to use a device with an unknown history.
Effects of resetting:
• All customer parameters are reset to their default values.
• Linearisation is switched to "linear", but the table values are kept. The table can be switched
back on in the "linearisation" (*4) function group in the "linearisation" (*40/*46) function.
In order to carry out the reset, enter "33333" in the "reset" (*A3) function in the "diagnostics"
(*A) function group.
Caution!
A reset leads to impairment of the measurement. A basic calibration is required after a reset.
Note!
The default values of each parameter are shown in bold in the menu overview in the appendix.