Diagnosis and troubleshooting

11.1

General troubleshooting

11.1.1
General errors
Device not responding
• Possible cause: Supply voltage does not match the specification on the nameplate
Remedial action: Apply the correct voltage
• Possible cause: The polarity of the supply voltage is wrong
Remedial action: Correct the polarity
• Possible cause: The cables do not contact the terminals properly
Remedial action: Check the electrical contact between cables and correct if necessary
• Possible cause: Load resistance too high
Remedial action: Increase the supply voltage to reach the minimum terminal voltage
No values visible on the display
• Possible cause: The plug of the display cable is not connected correctly
Remedial action: Connect the plug correctly
• Possible cause: Display is defective
Remedial action: Replace the display
"Communication error" is indicated on the display when the device is started or the
display is connected
• Possible cause: Electromagnetic interference influence
Remedial action: Check grounding of the device
• Possible cause: Defective cable connection or display plug
Remedial action: Replace the display
HART communication not working
• Possible cause: Communication resistor missing or incorrectly installed
Remedial action: Install the communication resistor (250 Ω) correctly
• Possible cause: Commubox is connected incorrectly
Remedial action: Connect Commubox correctly
Communication via CDI interface not working
Possible cause: Wrong setting of the COM port on the computer
Remedial action: Check the setting of the COM port on the computer and correct it if
necessary
11.1.2
Error - SmartBlue operation
Operation via SmartBlue is only possible on devices that have a display with Bluetooth
(optionally available).
Device is not visible in the live list
• Possible cause: Supply voltage too low
Remedial action: Increase the supply voltage.
• Possible cause: No Bluetooth connection available
Remedial action: Enable Bluetooth in the field device via display or software tool and/or
in the smartphone/tablet
• Possible cause: Bluetooth signal outside range
Remedial action: Reduce distance between field device and smartphone/tablet
The connection has a range of up to 25 m (82 ft)
• Possible cause: Geopositioning is not enabled on Android devices or is not permitted for
the SmartBlue app.
Remedial action: Enable/permit the geopositioning service on Android device for the
SmartBlue app.
Diagnosis and troubleshooting
49