Commissioning
Note!
This chapter describes the commissioning of the Gammapilot M using the operating module VU331
(which is located in the remote display and operating unit FHX40). Commissioning using the "ToF
Tool - FieldTool Package" is similar. For further instructions refer to the Operating Instructions of
the "ToF Tool - FieldTool Package" (BA224F).
6.1

Power up instrument

FMG60
V01.01.04 PA
Language
*92
ÂEnglish
Deutsch
Français
measured value
000
63.4 %
•••••••••••
Group selection
Âbasic setup
calibration
safety settings
After switching on the supply voltage, the instrument is first
initialised. Then, the following appear for approximately five
seconds:
• Device type: FMG60
• Software version: Vxx.yy.zz
– xx: hardware version
– yy: software version
– zz: software revision
• Type of the digital communicatioon signal:
– HART
– PA: PROFIBUS PA
– FF: Foundation Fieldbus
On first power-up you are requested to select the language for the
display texts.
Select the desired language with the O and S keys.
Confirm your selection by pressing F twice.
Now, the measured value display appears.
Press F in order to enter the group selection.
The "basic setup" function group is already marked.
Press F in order to enter the first function of the "basic setup"
function group.
Gammapilot M FMG 60 - PROFIBUS PA