Radiometric measurement
Hide thumbs Also See for PROFIBUS Gammapilot M FMG60:
Table of Contents

Advertisement

Accessories
"
1)
The weights of the various versions of the Gammapilot M are summarised in the section "Dimensions/Weight".
88
9.3.2
Application hints
The mounting device for level measurements and level limit detection consists of the bracket,
mounting clamps, retainers and the associated screws. The Gammapilot M is fitted into the bracket
by the centering knob and fixed to the tube by the mounting clamps and retainers. For a detector
length of 1600 mm or more two retainers are to be used. For the limit switch version (horizontal
mounting) the bracket is not needed.
A
(a)
(a)
(a)
(a)
(d)
(d)
A: A: A: A: level measurement, FMG60 without water cooling; B: B: B: B: level measurement, FMG60 with water cooling;
C: C: C: C: limit detection, FMG60 without water cooling; D: D: D: D: limit detection, FMG60 with water cooling;
E: E: E: E: horizontal mounting not allowed this way
(a): retainer for tube ∅ 80 mm; (b):
(b): retainer for tube ∅ 102 mm; (c):
(a):
(a):
(a):
(b):
(b):
(d): bracket
(d):
(d):
(d):
Caution!
• The mounting device must be installed in a way such that it can withstand the weight of the
1
Gammapilot M
under all operating conditions (e.g. vibrations).
• For a measuring length of 1600 mm or more, two retainers (vertical) or three retainers
(horizontally) must be used.
• For vertical mounting usage of the bracket is mandatory. Otherwise a sufficient stability and
support of the Gammapilot M is not ensured.
• For stability reasons the mounting version with the compartment housing pointing upwards
should only be used in exceptional cases(e.g. lack of space).
• To prevent damage of the detector tube, the maximum torque of the retainer screws is 6 Nm.
9.3.3
Product structure for complete mounting set
Application
1
Level, measuring range of FMG60: 400-1200 mm; (1 bracket, 1 retainer)
2
Level, measuring range of FMG60: 1600-2000 mm; (1 bracket, 2 retainers)
3
Limit switch, measuring range of FMG60: 200/400 mm; (2 retainers)
9
Special version, to be specified
Mounting clamp
A FMG60 without water cooling
B
FMG60 with water cooling
Y
Special version, to be specified
Material
1
316L
9
Special version, to be specified
FHG60 -
complete product designation
Gammapilot M FMG 60 - PROFIBUS PA
B
(c)
(c)
(c)
(c)
(d)
(d)
(c): retainer for water cooling jacket;, ∅140 mm;
(c):
(c):
C
(b)
(a)
D
(c)
(c)
E
(a)
(a)
L00-FMG60xxx-17-00-00-xx-008A