Endress+Hauser PROFIBUS Gammapilot M FMG60 Operating Instructions Manual page 74

Radiometric measurement
Hide thumbs Also See for PROFIBUS Gammapilot M FMG60:
Table of Contents

Advertisement

Commissioning
ENDRESS + HAUSER
+
E
ENDRESS + HAUSER
+
E
74
6.5.3
Calibration points
Excerpt from the operating menu
The following excerpt from the operating menu shows, how the density calibration points are
entered.
The individual functions are explained in the sections below.
The functions are only accessible after the background calibration has been performed.
*1A
*15
calibr. point
calibration
1 ... 9
0: stop/edit
point 10 for
1: start
recalibration
0
*1D
calibr. point
0: not used
1: used
2: clear
"Calibration point" (*1A)
calibr. point
*1A
Â1
2
3
This function is used to select, which calibration point will be entered.
Selection:
• "1" ... "9" : Calibration points for various densities
• "10": recalibration point
"Calibration" (*15)
calibration
*15
stop/edit
start
This function is used to start the automatic entering of the selected calibration point.
Selection:
• stop/edit
This option must be selected, if
– the calibration point is not to be entered (e.g. because it has already been entered). The pulse
rate of the calibration point is then displayed in the following function, "density calib." (*1B).
I frequired, this value can be changed.
– the calibration is to be entered manually. This can be done in the following function, "density
calib." (*1B).
• start
This option is used to start the automatic entering of the calibration point. After selecting this
option, the Gammapilot M changes to the "avg. pulse rate" (*11) function.
Gammapilot M FMG 60 - PROFIBUS PA
E
*11
*1B
avg. pulse
density calib.
1
rate
[c/s]
*1E
*1F
absorp. coeff.
ref. pulse rate
*1C
density value
Group
selection
*19
next
point
0: no
1: yes
L00-FMG60xxx-05-00-00-en-015