Gammapilot M FMG 60 - PROFIBUS PA
"
"
3.3

Water cooling

For the versions of the Gammapilot with water cooling jacket, the following applies:
• Material: 316L
• Water connection: 2 x G 1/4"A, DIN ISO 228
• Inlet temperature: max. 40 °C
• Outlet temperature: max. 50 °C (temperature monitoring recommended)
• Water pressure: 4 ... 6 bar
T < 75 °C
*
G1/4"A
Connections of the water cooling jacket
Caution!
• It is recommended to ground the water cooling jacket separately at the provided earth terminal
(see above picture)
• The ambient temperature of the compartment housing must not exceed 75 °C. This is also valid,
if water cooling is applied.
• The three screws, connecting the pipe hou-
sing to the compartment housing, must not be
opened.
3.3.1
Mounting versions
A
A: A: A: A: Recommended mounting position for level measurement: compartment housing at the bottom;
B: B: B: B: In exceptional cases (e.g. shortage of space) the compartment housing may be located at the top.
C: C: C: C: Mounting position for limit detection and density measurement.
Caution!
The water inlet must always be at the bottom to ensure that the water cooling jacket is completely
filled.
T < 120 °C
B
Installation
G1/4"A
L00-FMG60xxx-11-00-00-xx-001
L00-FMG60xxx-05-00-00-xx-052
C
L00-FMG60xxx-17-00-00-xx-009
19