Gammapilot M FMG 60 - PROFIBUS PA
!
4

Wiring

4.1

Terminal compartments

The Gammapilot M has got two terminal
compartments:
• terminal compartment A, for
– Power supply
– Signal output (depending on the instrument
version)
• terminal compartment B, for
– Signal output (depending on the instrument
version)
– PT-100 input (4-wire)
– Pulse input for cascading mode
– Pulse output for cascading mode
– Display and operating module FHX40 (or
VU331)
!
Note!
Depending on the instrument version, the signal
output is located in the terminal compartment A
or B.
4.2

Cable entry

The number and type of calbe entries depend on
the instrument version ordered. The following
types may occur:
• Gland M20x1,5
tightening diameter (clamping range):
7.0 ... 10.5 mm
• Thread M20x1,5
• Thread G1/2
• Thread NPT1/2
• Plug M12 (s. below "Fieldbus plug
connectors)
• Plug 7/8" (s. below "Fieldbus plug
connectors)
Additionally, the Gammapilot M has a socket for
connection of the separate display and operating
module FHX 40. The FHX 40 can be connected
without opening the hosuing of the
Gammapilot M.
Note!
On delivery, not more than one cable gland is present for each of the terminal compartments. If
further cable glands are required (e.g. for cascading mode), they must be supplied by the user.
B
A
B
1
A: A: A: A: Cable entries for terminal compartment A;
B: B: B: B: Cable entries for terminal compartment B;
1: 1: 1: 1: Socket for FHX 40
Wiring
L00-FMG60xxx-05-00-00-xx-041
A
L00-FMG60xxx-04-00-00-xx-003
21