Endress+Hauser PROFIBUS Gammapilot M FMG60 Operating Instructions Manual page 50

Radiometric measurement
Hide thumbs Also See for PROFIBUS Gammapilot M FMG60:
Table of Contents

Advertisement

Commissioning
50
6.2.1
Basic Setup
Step
Function
1
Present date (*01)
2
Isotope (*03)
3
Operating mode (*04)
4
Measuring mode (*05)
the following functions are valid for density and concentration measurements only:
5
Density unit (*06)
6
Min. density (*07)
(lower limit of measuring range)
7
Max. density (*08)
(upper limit of measuring range)
8
Pipe diameter unit (*09)
9
Pipe diameter(*0A)
the following function is valid for all measuring modes:
10
Output damping (*0B)
Gammapilot M FMG 60 - PROFIBUS PA
Input or selection Remarks
Adjust day by O and S ;
e.g
confirm by F;
12.11.04 10:30
same procedure for month, year, hour
and minute
Select by O and S;
137
Cs
Confirm by F
60
Co
• no compensation
Select by O and S;
• stand alone
Confirm by F
• master
• slave
• end slave
• not defined
Select by O and S;
• level
Confirm by F
• limit
• density
• concentration
Select by O and S;
3
• g/cm
Confirm by F
• g/l
• ...
Adjust each digit by O and S and
e.g.
confirm by F
3
0,9500 g/cm
after the last digit: confirm the
complete value by F.
Adjust each digit by O and S and
e.g.
confirm by F
3
1,2500 g/cm
after the last digit: confirm the
complete value by F.
Select by O and S;
• mm
Confirm by F
• inch
Adjust each digit by O and S and
e.g.
confirm by F
200 mm
after the last digit: confirm the
complete value by F.
Adjust each digit by O and S and
e.g. 60 s
confirm by F
after the last digit: confirm the
complete value by F.
see section
6.3.1
6.3.2
6.3.3
6.3.4
6.3.5
6.3.6
6.3.7
6.3.8
6.3.9
6.3.10

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hart gammapilot m fmg60

Table of Contents