Endress+Hauser PROFIBUS Gammapilot M FMG60 Operating Instructions Manual page 80

Radiometric measurement
Hide thumbs Also See for PROFIBUS Gammapilot M FMG60:
Table of Contents

Advertisement

Troubleshooting
80
Code
Description
A633
empty/free not calibrated
A634
density not calibrated
A635
present date not defined
A636
calibration date not plausible
A637
operating mode not defined
A638
measurement mode not defined
A639
temperature compensation not complete
W662
high sensor temperature (warning)
A663
sensor temperature too high (alarm)
A664
temperature measurement error
W681
current out of range
A692
gammagraphy detected (alarm)
W693
gammagraphy detected (warning)
Gammapilot M FMG 60 - PROFIBUS PA
Remedy
perform empty/free calibration
("Abgleich" (*1) function group)
• Check: Has at least one calibration point been entered and activa-
ted?
If not: Enter and activate calibration point(s). ("calibration" (*1)
function group)
• Check: Is the "reference pulse rate" (*1F) larger than 2
If yes: re-perform density calibration ("calibration" (*1)) func-
tion group
enter present date
("basic setup" (*0) function group
"todays date" (*01) function)
check calibration date and enter it again
("system parameters" (*C) function group
"calibration date" (*C7) function)
enter operating mode
("basic setup" (*0) function group
"operating mode" (*04) function)
enter measurement mode
("basic setup" (*0) function group
"measurement mode" (*05) function)
enter at least two value paris "temperature density"
"temperature compensation" (*3) function group)
install water cooling jacket or thermal screening
install water cooling jacket or thermal screening
check correct functioning and connection of the PT-100 sensor
check calibration and linearisation
• Check, if interference radiation is present or the "hold time"
(*54) is too short.
• If no interference radiation is present: decrease gammagraphy
sensitivity
("gammagraphy" (*5) function group
"sensitivity" (*52) function)
wait for end of the gammagraphy measurement
32
?

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hart gammapilot m fmg60

Table of Contents