Chapter 6 Method of Use
(3) Turn the connector of water tank attached to the endoscope
to remove it.
Then cover the feed water connector of the endoscope
with a finger.
(4)
Continue depressing the air/water button until the water
remaining in the channel is discharged completely.
6.5.3

Pre-cleaning Suction Channel

(1) Immerse the distal end of endoscope in cleaning fluid,
and depress suction button to suck cleaning fluid for 10
seconds.
(2)
Keeping suction button depressed, pull the distal end of
endoscope out of cleaning fluid and suck in air.
(3) Repeat steps (1) and (2) above 2 or 3 times to suck
cleaning fluid and air alternately.
(4)
Finally, pull distal end of endoscope out of cleaning fluid
and suck in air until cleaning fluid is completely drained
from suction channel.
6-22
Remove body fluids from the suction channel of endoscope in accordance with the
following procedure.