Chapter 6 Method of Use
(3) Unlock the bending portion by turning the up-down
locking lever and the left-right locking knob in the
direction of F
[Note]
Another procedure is also available: you can insert the
endoscope by locking the bending portion only in the
left-right direction and unlocking it in the up-down
direction.
(4) Switch on the power to the processor and turn on the lamp.
(5) Deflate the balloon on the distal end of endoscope and the
balloon on the Over-tube.
Apply clean lubricant (Xylocaine jelly or the like) to the
insertion portion as required.
[Note]
Do not apply Xylocaine spray, olive oil or the like directly
to the insertion portion.
(6) Insert the distal end of the endoscope from the oral cavity
to the pharynx, while observing the process.
Adjust the brightness with the level button on the light
source.
6-6
until they stop.