Problem
A live image is
not displayed after
image freezing is
cancelled during an
examination.
An image is suddenly
discolored during an
examination.
An image appears
distorted.
Air and/or water
cannot be fed.
Low air/water supply
volume
[Note] Reset: Turn off the processor and the light source, and wait for at least 5 seconds. Turn on the processor and the
light source again, and then light the lamp by pressing the Lamp button.
Cause
The system has malfunctioned due to such as
static charges.
1) The system has malfunctioned due to such
as static charges.
2) The video signal cable has shorted out or
broken.
3) The imaging section is damaged.
1) High-frequency interference
2) Not connected correctly
3) The video signal cable has shorted out or
broken.
4) The imaging section is damaged.
1) The pump is switched off.
2) The water tank cap is loose.
3) The water tank is filled with too much
water.
4) The water tank is empty.
5) The water tank is not connected.
1) Foreign matter adheres to the air/water
supply channel.
2) The air/water supply channel is damaged.
Remedy
Reset the processor and the light source.
If the image is still not displayed, turn the
processor and the light source off, and then
straighten the bending portion to unlock
and release the angle knobs. Pull out the
endoscope slowly.
1) 2) 3)
Reset the processor and the light source.
If the image is not recovered and it is
impossible to continue the examination, turn
the processor and the light source off, and
then straighten the bending portion to unlock
and release the angle knobs. Pull out the
endoscope slowly.
1) Stop power supply to the diathermic
treatment equipment to restore image
output.
The endoscope is working all right.
2) Connect properly.
3) 4)
Reset the processor and the light source.
If the image is not recovered, turn the
processor and the light source off, and then
straighten the bending portion to unlock
and release the angle knobs. Pull out the
endoscope slowly.
1) Switch on the pump.
2) Close the cap firmly.
3) Reduce the water level in the water tank to
about 80% of its capacity.
4) Fill the water tank with water.
5) Connect the water tank.
1) Pull out the endoscope from the body cavity
and clean the air/water supply channel
referring to "6.5.1 Wiping off Endoscope"
and "6.5.2 Purging Air/Water Supply
Channel."
If the air/water supply volume is still low,
replace the endoscope with a spare one.
2) Consult your local FUJIFILM dealer.
Appendix
Appendix-9