6.5.4
Cleaning Air/Water Supply Channel Using CA-511N
This section explains how to clean the air/water supply channel using the CA-511N (optional
item).
For details on how to use the CA-511N, refer to the operation manual of the CA-511N.
(1)
Perform pre-cleaning of items other than the air/water
supply channel referring to 6.5.1 "Wiping off Endoscope"
and 6.5.3 "Pre-cleaning Suction Channel".
(2)
Immerse the distal end of endoscope in clean water.
(3)
Turn OFF the air supply pump of the light source.
(4)
Remove the air/water button (AW-500) from the
endoscope and then immerse it in cleaning fluid.
(5) Attach the cleaning adapter CA-511N to the air/water
supply valve of the endoscope.
(6)
Turn ON the air supply pump of the light source.
(7)
Press the button on the cleaning adapter CA-511N to
supply water for 30 seconds.
Make sure that water discharges from the distal end of
endoscope while pressing the button.
(8)
Release the button to supply air for 10 seconds or longer
until water in the air/water supply channel is completely
discharged.
Chapter 6 Method of Use
6-23