Chapter 5 Preparation for Use of Endoscope
(3) Feed air from the balloon air feed inlet several times and
let air come out from the balloon air feed outlet.
[Note]
Discharge water completely to remove moisture from
inside the channel.
5.7.2
Mounting Endoscope Balloon
[Note]
Use the balloon (BS-1 or BS-2) applicable to this product.
(1)
Prepare the balloon, setting tools (ST-01B and ST-05B)
and fixing rubber.
ST-01B : Fixing rubber setting tool
ST-05B : Balloon setting tool
[Note]
Do not use the balloon after its expiration date has elapsed.
(2) Check that the tip of the endoscope is free of dirt.
Apply a small amount of 70% ethanol solution to the tip
of the endoscope.
5-18
ST-05B
Fixing rubber
ST-01B
Balloon