Chapter 5 Preparation for Use of Endoscope
(3) Connect the tube 1 (white tube) of the balloon controller
to the air feed inlet (clear tube) of the Over-tube.
5.9.2
Inspecting Balloon (Balloon Controller)
(1)
Make sure that the "SCOPE" side connector of the balloon
controller is connected to the balloon air feed inlet of the
endoscope, and the "OVERTUBE" side connector of the
balloon controller is connected to the air feed inlet of the
Over-tube.
(2) Press the endoscope-side switch on the balloon controller,
and then make sure that the balloon attached to the distal
end of endoscope inflates.
(3)
Press the switch again, and then make sure that the balloon
deflates.
(4)
Press the Over-tube-side switch on the balloon controller,
and then make sure that the balloon attached to the distal
end of Over-tube inflates.
(5)
Press the switch again, and then make sure that the balloon
deflates.
5-26