Electromagnetic immunity compliance information and guidance
Immunity test
Electrostatic discharge (ESD)
EN 61000-4-2
Electrical fast transient/burst
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN 61000-4-11
Power frequency (50/60 Hz)
magnetic field
EN 61000-4-8
Appendix-6
EN 60601-1-2
Test level
± 6kV: contact
± 8kV: air
± 2kV: for power supply lines
± 1kV: for input/output lines
± 1kV: Line to line
± 2kV: Line to earth
< 11.5V (> 218.5V dip)
For 0.5 cycle
92V (138V dip)
For 5 cycle
161V (69V dip)
For 25 cycle
< 11.5V (> 218.5V dip)
For 5 sec
3 A/m
Compliance
level
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
Same as left
synthetic material, the relative humidity
should be at least 30%.
Main power quality should be that of a
Same as left
typical commercial or hospital.
Main power quality should be that of a
Same as left
typical commercial or hospital.
Main power quality should be that of a
typical commercial or hospital. If the
user of this product requires continued
Same as left
operation during power mains
interruptions, it is recommended
that this product is powered from an
uninterruptible power supply or battery.
It is recommended to use this product
by maintaining enough distance from
Same as left
any equipment that operates with high
current.
Guidance