Required Tools; Power Button Pca Replacement; Required Parts; Power Button Pca Removal - FujiFilm SonoSite sII Service Manual

Sonosite sii
Hide thumbs Also See for SonoSite sII:
Table of Contents

Advertisement

Caution:
Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable
at first. Initial ESD symptoms may be slightly degraded performance or image quality.

Required Tools

• #1 Phillips screwdriver
• 1/4 inch flat blade screwdriver
• Torque screwdriver, 2.0–15.0 inch pounds (0.23–1.7 newton-meter)
• Anti-static mat
• Wrist grounding strap

Power Button PCA Replacement

Required Parts

• P21439, Power Button PCA

Power Button PCA Removal

Note: The User Interface does not need to be removed to replace the Power Button PCA. Only the Top
Cap assembly should be removed to replace this item.
1 Remove Top Cap Assembly referenced in
2 Remove two screws from Power PCA shown in
Remove
two screws
Figure 6.9 Power Button PCA
3 Remove PCA far enough to disconnect Power Cable. If cable is not damaged, it can be reused.
Power Button PCA Replacement
1 Connect Power PCA cable to Power Button PCA and place as shown in
2 Install two screws and torque to 3.0 in-lbs.
3 Install Top Cap Assembly by performing steps 8-11 in reverse order in
on page 14.
44
Chapter 6: Replacement Procedures
"User Interface Removal"
on page 37, steps 1 - 4.
Figure 6.9
Figure 6.9
"Install User Interface"
on page 42

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents