Reference Level Accuracy; Test Limits; Test Description - Agilent Technologies ESA-E Series User Manual

Esa spectrum analyzers, 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

Functional Testing

Reference Level Accuracy

Reference Level Accuracy

Test Limits

Reference Level
dBm
dBmV
–30
21.76
–20
31.76
–10
41.76
–40
11.76
–50
1.76
–60
–8.24
–70
–18.24

Test Description

A 50 MHz CW signal is applied to the Input of the analyzer. The
amplitude of the source and the analyzer's reference level are decreased
in 10 dB steps. The analyzer marker functions are used to measure the
amplitude difference between steps. Reference Level Accuracy is tested
in both Log and Linear Scale Modes. Most of the error is contributed
from the output attenuator inaccuracy in the synthesized sweeper and
not the analyzer.
3-32
Minimum
Maximum
(dB)
(dB)
Reference
Reference
–1.40
1.40
–1.40
1.40
–1.40
1.40
–1.40
1.40
–1.40
1.40
–2.0
2.0
Chapter 3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents