Procedure (12.01 Ghz To 13.2 Ghz) E4405B - Agilent Technologies ESA-E Series User Manual

Esa spectrum analyzers, 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

4. Press the following keys on the analyzer:
,
Sweep
Sweep (Cont)
BW/Avg, Average (Off)
,
BW/Avg
Resolution BW (Auto)
Video BW (Auto)
,
SPAN
50 kHz
FREQUENCY
5. Press
. Set the center frequency of the analyzer to the
Center Freq
frequency recorded in the Measured Frequency column (i) in
Table 3-1
for 6.01 GHz to 12.0 GHz.
6. Press the following keys on the analyzer:
,
BW/Avg
1 kHz
Video BW Auto Man
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
,
Display
Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 12 in
Table
3-1. The average noise level should be less than the Maximum.

Procedure (12.01 GHz to 13.2 GHz) E4405B

1. Press
Preset System
analyzer. Wait for the preset routine to finish. Set the analyzer by
pressing the following keys:
,
FREQUENCY
Start Freq
,
Stop Freq
13.2 GHz
,
AMPLITUDE
–70 dBm
,
Attenuation (Man)
,
BW/Avg
1 MHz
Video BW Auto Man
Chapter 3
,
30 Hz
,
,
Power On/Preset
Preset (Factory)
,
12.01 GHz
0 dB
,
10 kHz
Functional Testing
Displayed Average Noise Level
,
, on the
Preset
3-17