Pump Malfunction - EBARA EV-A Series Instruction Manual

Dry vacuum pump
Table of Contents

Advertisement

8.2 Pump Malfunction

When abnormality occurs in the pump, the following message is displayed in the LED display on the
controller.
Press STOP/RESET button on the controller after solving the problem.
be reset.
LED
display
E01.□
During low speed
E02.□
During deceleration
E03.□
During acceleration
E04.□
Other
Overload protection
E05.□
E07.□
Over voltage protection
E08.□
EEPROM error
E09.□
Under voltage
E10.□
Current detected error
E11.□
CPU error
E14.□
Earth protection
E15.□
Incoming high voltage protection
E19.□
Temperature detection error
E21.□
Fin temperature error
E22.□
CPU communication error
E25.□
Main circuit error
E30.□
Driver error
E35.□
Thermistor error
E37.□
Thermostat error
E38.□
Overload protection during low speed
E50.□
Shaft lock
E51.□
Low speed
Note 1)
:Abnormality will not be reset for 10 seconds after alarm occurred.
How to use LED monitor
PM10U
Contents
High voltage
Highly loaded running
protection
Backing pressure of the pump is abnormal
・ Sudden load shift, voltage in power uprising
Internal memory error
Power voltage abnormality
Inverter error
Pump error
Inverter error
Pump error
Pump temperature uprising
Overload when the pump starts
Overload when the pump starts
Number of rotation is decreasing
Cause
LED Display
UP Switch
DOWN Switch
Fig. 8.1 Pump operation side
The error will
Corrective measures
Needs to overhaul pump.
(Foreign object may intrude into pumping room.)
Check outlet piping.
Check voltage in power is uprising.
Needs to check and repair of electric parts.
Check voltage in power is short.
Needs to check and repair electric parts.
Needs to overhaul pump.
There should be source of noise near the pump.
Consider plan to decrease its noise
Needs to check and repair electric parts.
Needs to overhaul pump.
Make sure surrounding temperature is proper
range.
Needs to check and repair electric parts.
(air-cooled fan may be stopped)
Make sure outlet piping is not stuffed.
Needs to overhaul pump.
(Foreign object may intrude into pumping room.)
PROGRAM Lamp
STOP/RESET Switch
EBARA
P.45
CORPORATION