RIGOL
Aux Output
You can set the type of the signal output from the [TRIG OUT] connector on the
rear panel.
Press Utility  System, and then press the AUX Out key continuously to select
"TrigOut" or "PassFail".
Trigger Output
After this type is selected, at each trigger (hardware trigger), the oscilloscope
outputs a signal from the [TRIG OUT] connector on the rear panel that can
reflect the current capture rate of the oscilloscope. Connect this signal to a
waveform display device, measure the signal frequency. It is found that the
measurement result is the same as the current capture rate.
Pass/Fail
When this type is selected, the instrument can output a positive or negative
pulse via the [TRIG OUT] connector when a successful or failed event is
detected. Refer to descriptions in "To Set the Output Form of the Test
Results". When you enable the Aux output, the AUX Out menu is
automatically set to "PassFail". When the AUX Out menu is automatically set to
"TrigOut", then in the pass/fail test menu, the Aux Out function is automatically
disabled.
Help
Press Utility  System  Help to enter the "Help" function menu. You can also
enable the touch screen and then tap the function navigation icon
lower-left corner of the screen to open the function navigation. Then, tap the "Help"
icon to open the "Help" function menu.
Press About to view the system information of the oscilloscope.
Press Help to enter the built-in help system.
Press Option list to view the name of the installed option and other detailed
information about the option from the option list.
Press Option install to install the desired option. For detailed installation
procedures, refer to descriptions in "To View the Option Information and
the Option Installation".
Press Online upgrade to perform online upgrading for the system software.
Press Local upgrade to perform local upgrading for the system software.
19-8