Chapter 5 To Trigger the Oscilloscope
<>: triggers when the runt pulse width is greater than the lower limit and
smaller than the upper limit of pulse width. Press Upper to set the maximum
pulse width of Runt trigger. Press Lower to set the minimum pulse width of

Runt trigger.

Note:
Level Selection and Adjustment of Trigger Level:
After the trigger condition setting is completed, you need to adjust the trigger level
to correctly trigger the signal and obtain a stable waveform. Press Level Select
(when "<>" is selected as the trigger condition, Level Select is a sub-menu under
More), then rotate the multifunction knob
adjustment, and then press down the knob to select the type. You can also press
Level Select continuously to select the level type or enable the touch screen to tap
the desired level type and select it.
Level A: only adjusts the upper limit of the trigger level, and the lower limit of the
trigger level remains unchanged.
Level B: only adjusts the lower limit of the trigger level, and the upper limit of the
trigger level remains unchanged.
Level AB: adjusts the upper and lower limits of the trigger level at the same time, and
the trigger level deviation (the difference between the upper limit and lower limit of
the trigger level) remains unchanged.
Note:
level adjustment type.
Rotate the Trigger
During the adjustment of trigger level, two trigger level lines appear on the screen,
and they move up and down with the change of the trigger level. At the same time,
the real-time trigger level information is displayed at the lower-left corner of the
screen (as shown in the figure below, H indicates the upper limit of the trigger level,
L indicates the lower limit of the trigger level, and △ indicates the trigger level
deviation). When you stopping modifying the trigger level, the trigger level line and
the trigger level information at the lower-left corner of the screen disappear in about
2 s. The current trigger level deviation is displayed at the upper-right corner of the
screen.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch
the current trigger mode. For details, refer to descriptions in "Trigger Mode".
Trigger Parameter Setting:
Set the trigger parameters (trigger holdoff and noise rejection) under this trigger
MSO5000 User's Guide
nd Quality Products Online at:
The lower limit of the pulse width must be smaller than the upper limit.
In Runt trigger, pressing Trigger
LEVEL knob to adjust the corresponding trigger level.