Infusomat® Space 6.0
Unit Diagnosis / Calibration
13. If the safety clamp (locking clamp or clamping lever ) open,
you are requested to close the safety clamp. Close the safety
clamp with the slide clamp.
The "Lock bolt calibration" group box is now activated. The
lock bolt is moved into the open position.
14. Check the position of the lock bolt in the device. The lock bolt
lugs must not be visible in the three openings. It must be pos-
sible to tilt the operating unit almost as far as the closed po-
sition (gap between upper edge of the operating unit and
upper edge of the housing top is about 5 mm).
15. Check the gap and confirm the lock bolt position by pressing
the "Yes" button.
The lock bolt automatically closes.
16. Check the position of the lock bolt in the device. The lock bolt
lugs are visible in the three openings. It must not be possible
to tilt the operating unit as far as the closed position (gap be-
tween upper edge of the operating unit and upper edge of the
housing top is about 10 mm).
17. Check the gap and confirm the lock bolt position by pressing
the "Yes" button. The lock bolt is moved into the open posi-
tion.
18. After calibration you are prompted to switch the device off.
Disconnect all connectors from the device.
19. Switch the device off (without connection to the power sup-
ply) and press the "OK" button.