Error Code
1
2001 ... 2013
Internal Error
2
2014 ... 2016
Loudspeaker
3
2017
KUP does not work
4
2018
FUP-KUP OS Cycle Timeout
5
2019
FUP Flash Memory Error
6
2020
FUP different version KuP to FuP Software defective or obsolete
7
2021
FUP UPSTREAM TOO HIGH
8
2022
FUP UPSTREAM TOO LOW
9
2023
FUP DOWNSTREAM TOO HIGH
10
2024
FUP DOWNSTREAM TOO LOW
11
2025
FUP EA key closed too long
12
2026
Internal Error
13
2027 ... 2033
FUP Airsensor
FUP Temperature Sensors
14
2034
FUP CAN packet scheduler list
failure
15
2035
Light barrier or motor defective/
turn backwards
16
2036
FUP Language Flash Memory Er-
ror
17
2037
FUP Language Version Error
18
2038
Internal Error
19
2039
FuP collective CAN Alarm
20
2040 ... 2041
Internal error
21
2042 ... 2051
FUP Syncdata KuP
22
2052 ... 2053
Internal error
23
2060 ... 2085
TCI Error
24
2088 ... 2090
Internal error
Table 2 - 2 Error Codes of the Function Processor
EN
Infusomat® Space 6.0
Definition
Loudspeaker connector
Loudspeaker
Software defective or obsolete
Software defective or obsolete
Software defective or obsolete
Upstream pressure sensor defective Replace upstream pressure sensor
Upstream pressure sensor defective
Downstream pressure sensor de-
fective
Downstream pressure sensor de-
fective
Keyboard defective
Air sensor defective
CAN communication error
Light barrier defective
Language file faulty
Language file faulty
Two pumps with identical serial
number in system
Error in synchronization with the
control processor
Error in relation to TCI therapy to-
gether with special settings
Unit Diagnosis / Calibration
Error Codes of the Function Processor
Possible Cause
Check the loudspeaker connector.
Check the loudspeaker.
Update software, change ISP processor
PCB if necessary
Replace downstream pressure sensor
Exchange operating unit
Check ribbon cable of air sensor.
Replace air sensor
Restart device
Exchange processor PCB
Update software
Update software
Check and correct serial numbers.
Restart the device, change settings
slightly if necessary. If the error persists:
Please load the history file via HiBaSeD
and inform B. Braun
2
Fault Clearance
2 - 3