Braun Infusomat Space Service Manual page 39

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

History Entry
CAN Error Code
1
CAN Error Code
2
Table 2 - 5
EN
Infusomat® Space 6.0
Value
Duplicate serial number in the system Calibrate the device using HiBaSeD and check
Duplicate address assigned in the sys-
tem
Unit Diagnosis / Calibration
c) If the fault does not occur with the first two pumps, test
the other pumps affected by the communication problem
on the same slot until you have found the defective
pump.
Eliminating the CAN-Dead state
1. Connect the HiBaSeD in order to eliminate the CAN-Dead
state.
2. Carry out calibration.
3. Take particular care to enter the serial number correctly.
Entering the states in the history
In order to provide diagnostic information to service technicians,
the various CAN states are entered in the history.
They can then recognise, for example, whether there was a dupli-
cate INR in the system or whether address assignment function of
a Station is defective.
The following history entry has to be evaluated for this purpose:
Meaning
the serial number in the process.
Check the SpaceStation
Address assignment defective
2
Measure
2 - 9