Braun Infusomat Space Service Manual page 123

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

7 - 2
Test Equipment and Special Tools
7 -
Test equipment
Fig.: 7 - 1 Slide clamp
EN
Infusomat® Space 6.0
Designation
For Calibration
Space service connector . . . . . . . . . . . . . . . . . . . . . . . . 3452 1062
HiBaSeD Service-CD. . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3301
HiBaSeD Service-CD (US version) . . . . . . . . . . . . . On request
Infusomat® Space Line, standard type
(new, not used)
Infusion bag or bottle, min. 100 ml
Three-way valve (2 pieces)
10 ml syringe (air buffer for measurement with
electronic pressure gauge)
(syringe drawn up to 10 ml and piston fixed mechanically)
1 ml syringe for air bubble injection
Electronic pressure gauge with peak value recognition (accuracy
class 1)
Pressure cell (alternative of electronic pressure gauge)
Slide clamp (of Infusomat® Space Line)
For Repairs
HiBaSeD Service-CD. . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3301
HiBaSeD Service-CD (US version) . . . . . . . . . . . . . On request
For the TSC
Measuring instrument for electrical safety
Service adapter SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0770 5174
for measuring the electrical safety
Infusomat® Space Line, standard type (new, not used)
Infusion bag or bottle, min. 100 ml
Three-way valve (2 pieces)
10 ml syringe (air buffer for measurement with
electronic pressure gauge)
(syringe drawn up to 10 ml and piston fixed mechanically)
1 ml syringe for air bubble injection
Electronic pressure gauge with peak value recognition (accuracy
class 1)
Pressure cell (alternative of electronic pressure gauge)
Graduated cylinder 25 ml, +/- 0.4 ml
Stop watch
7
Order No.
7 - 1