3
Disassembly / Assembly
Fig.: 3 - 45
3 - 44
Inner Frame and Housing Front Panel
1. Move the lock bolt by hand to the left until stop (open posi-
tion) before mounting the housing front panel.
2. Close the safety clamp with a slide clamp after installation.
3. Push the membrane carefully over the pump drive pusher.
4. Route the cable of the lock bolt drive in the cable guide on
the lock bolt drive and between the two cable guide pins of
the front frame.
Note
Always use a new membrane for assembly. Use Meliseptol® to fa-
cilitate installation of the membrane over the pushers.
5. Engage the front frame lugs (arrow,
the inner frame openings.
Note
When mounting the inner frame and the housing front panel,
make sure that the operating unit is opened.
Fig.: 3 -
45) exactly in
Infusomat® Space 6.0