4 - 4
Cleaning and Disinfecting
Fig.: 4 - 1
EN
Infusomat® Space 6.0
Servicing the Unit
4 -
Notes
WARNING
!
Liquid may penetrate live parts of the unit.
DANGER OF ELECTRIC SHOCK.
-
Before disinfecting the pump, always disconnect the pump
from Infusomat® Space the patient, switch off the device and
disconnect from power and other devices (e.g. staff call).
-
To reduce the likelihood of moisture ingress into the electrical
connectors, the P2 connector of a power supply or connec-
tion cable may be used to cover the connections during
cleaning operations. Ensure that any connectors used to cov-
er are not connected to a wall outlet or other voltage source.
-
Do not touch any parts of the wiring or the tube pump with
sharp objects.
-
We recommend using the cleaning and disinfecting agents
Meliseptol® Wipes sensitive or Meliseptol® Foam pure from
B. Braun for cleaning and disinfecting.
-
In the case of severe soiling that cannot be removed by the
above method, please contact the hospital's technical de-
partment.
-
The current internal hygiene measures of the hospital apply.
Housing
Clean and disinfect all surfaces of the Infusomat® Space.
Start by cleaning the display with a dampened nonwoven cloth.
Clean using a damp and NOT a soaking cloth. Dispose of the cloth
after use.
We recommend: Meliseptol® Wipes sensitive or Meliseptol®
Foam pure.
You can use Meliseptol® Foam pure to clean the flex cable of the
Infusomat® Space. Apply with a cotton swab..
Make sure that no liquid is allowed to enter the contacts or ven-
tilation slots. Allow the device to dry after cleaning/disinfecting
(for at least 1 minute).
When they are dry, insert the cleaned/disinfected Space infusion
pumps back into the holder or SpaceStation (see cleaning plan for
SpaceStation) and protect from dirt until they are next used.