Fig.: 2 - 1
EN
Infusomat® Space 6.0
Unit Diagnosis / Calibration
Note
Please note that text and / or functions of the Service Program
may change depending on the software version. The following
screenshots are only examples and represent the state when the
manual was printed.
1. Start the Service Program
➨ pg. 1 -
10).
2. Select the tab "Calibration".
3. Press the "New device" or "Modify" command button in the
"Calibration procedure" group box (depending on the desired
calibration scope). The device is switched on.
Note
"New device" initiates a complete calibration. With "Modify", you
can select the desired calibration steps yourself.
(see „Starting the Service Program"
2
2 - 11