Visual Inspection
M686 70 01 05 F04 / 38918838
5 - 2
Technical Safety Check (TSC)
5 -
Electrical Safety
According to IEC 60601-1 or IEC 62353
(Master – to be added to the documentation)
Functional Inspection
Operation (continued)
❒ Trigger bolus at the device
❒ Trigger bolus by pressing the PCA but-
ton
❒ Slide clamp is completely pulled on to
the Infusomat® Space Line when the
operating unit is opened.
❒ The operating lever of the safety clamp
can be pressed down uniformly against
spring tension and engages. The yellow
caution sign flashes.
Pressure cut-off, electronic:
Delivery rate: 250 ml/h
Test without drop sensor
❒ Upstream
❒ Alarm
❒ Downstream
❒ Pressure stage 2
(0,1 ... 0,7 bar)
❒ Pressure stage 5
(0,4 ... 1,0 bar
❒ Pressure stage 9
(0,8 ... 1,4 bar)
Pressure limitation, mechanical:
❒ Pmax (1,8 ... 2,5 bar)
❒ Pmin (> 1,5 bar)
Safety clamp
❒ Pmin (> 1 bar)
Delivery accuracy
Ambient temperature 20 ... 25°C,
Delivery rate: 250 ml/h, Measured volume: 25 ml
❒ Divergence (± 5 %)
Drop sensor (optional)
Delivery rate: 250 ml/h
❒ Drop sensor alarm, flow
❒ Drop sensor alarm, occlusion
Infusomat® Space 6.0
Index h
____ bar
____ bar
____ bar
____ bar
____ bar
____ bar
____ %
Sheet 2 of 3
EN