Braun Infusomat Space Service Manual page 116

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

6
Procedural Instructions on the TSC
Fig.: 6 - 4
6 - 4
2. Battery check
Perform the battery service to discharge the battery until the
battery pre-alarm is issued (no bar displayed for battery dis-
charge level
(Fig.: 6 -
4), but it must not flash).
Open and close the operating unit 5x.
-
Replace the battery if the battery alarm sounds while you
open or close the operating unit.
3. Switch on the device with the power supply connected and
check the following points:
-
Self-test
Display on the LC display
-
Indicator lamps
LEDs (yellow, green, blue)
LED changes from green to red
-
Audible alarm
A low and then a high sound
-
Visual alarm
Red LED lights up for a short moment
-
Self-test
Display on the LC display
-
Status display
Battery capacity, mains operation
-
Lighting
LC display lighting, keyboard lighting
4. Operation
a) Open operating unit.
Yellow warning light beside the safety clamp flashes.
b) Insert Infusomat® Space Line.
Slide clamp releases release lever of the safety clamp.
The Infusomat® Space Line is pinched.
c) Close operating unit.
*
As of software version J: yellow, red, blue
As of software version J: from red to green
*
light up for a short moment
Infusomat® Space 6.0
EN

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents