Braun Infusomat Space Service Manual page 64

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

3
Disassembly / Assembly
Fig.: 3 - 17
Legend of fig. 3 - 17:
Item Designation
1
Removal aid
2
Fixation
3 - 18
6. If the retrofitting kit is used, replace the battery slide on the
battery module (nominal dimension after replacement: 2.15
mm ±0.15 mm).
7. Clean the bonding areas with isopropanol and let them air.
8. Position the removal aid
tion
(Fig.: 3 - 17 / Item
9. Press the removal aid onto the battery bottom in order to
1
avoid bubbling during bonding.
2
10. After retrofitting, reinsert the battery module and place on
the battery compartment cover.
11. Carry out a function test (in battery operation).
-
Set the max. pump rate (duration of pump operation ap-
prox. 1 min). During pump operation simulate a vibratory
load on the pump (shaking).
(Fig.: 3 - 17 / Item
1) over the fixa-
2) and move it.
Infusomat® Space 6.0
EN