Infusomat® Space 6.0
Procedural Instructions on the TSC
d) Staff call
- Plug service connector Space onto connector "P2".
- Lock roller clamp of Infusomat® Space Line.
The red LED on the Space service connector lights up
after an alarm is issued.
e) Carry out infusion with intermediate bolus and press all
buttons at least once.
Infusion and bolus are performed and all buttons trigger
the desired function.
f)
Trigger the bolus administration with the PCA button.
For this purpose, select a PCA drug from the database,
place the PCA button on P2 and trigger the bolus.
Ensure that the green LED at the PCA button lights up.
The bolus is transported.
g) Open operating unit.
Slide clamp of the Infusomat® Space Line is completely
pulled on to the line.
h) Operating lever of the safety clamp can be pressed down
easily against spring tension, engages and the caution
sign flashes yellow.
*
As of software version L07, the service connector is only detected if it is plugged
in before the pump is ready to start, for example, before inserting a disposable ar-
ticle
6
*
6 - 5