Error Code
28
2180 ... 2184 DRUG Error
DIFF Error
29
2199 ... 2200 Internal Error
30
2201
different version FuP to KuP
31
2202
Stepper motor PH1A not 0
32
2203
Stepper motor not low current
33
2204
Stepper motor K_SM_CLK
34
2205
Stepper motor PH1A not 1
35
2206
Stepper motor not zero current
36
2207
Stepper motor not middle current
37
2208
Stepper motor not high current
38
2209
Stepper motor PH1A, PH1B, PH2A,
PH2B not 0
39
2210
K_PKM_UB not off
40
2211
K_PKM_UB not on
41
2215
no V_MOT
42
2216
Overvoltage test failure
43
2217
no V_MOT
44
2218
Undervoltage test failure
45
2220
Door potentiometer defective
46
2221
Locking bolt position failure
47
2222
Temperature sensor difference
48
2223
Air sensor defective
49
2224
Rate compensation difference
50
2225
Rate compensation max value
51
2226
Rate compensation min value
52
2227
Received air data defective
53
2228
Temperature sensor switch defective
Table 2 - 3 Error Codes of the Control Microprocessor (Part 2 of 3)
EN
Infusomat® Space 6.0
Definition
Error in relation to dose calcula-
tion together with special set-
tings
Software
Stepper motor defective
Lock bolt drive defective
Processor PCB defective
Lock bolt drive (potentiometer)
defective
Software defective or obsolete
Air sensor defective
Disposable articles table for
Space Line Neutrapur incompat-
ible with existing air sensor
Air sensor defective
Rate compensation error
Air sensor defective
Unit Diagnosis / Calibration
Possible Cause
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Update unit software
Replace pump drive motor or proces-
sor PCB
Replace lock bolt drive or processor
PCB
Replace processor PCB
Replace lock bolt drive
Perform calibration.
Update software, change ISP proces-
sor PCB if necessary
Replace air sensor
Compare air sensor version with dis-
posable article table and import new
disposable article table, if necessary.
(see „Rework when Replacing the Air
Inline Sensor" ➨ pg. 3 - 52)
Replace air sensor
Update unit software
Replace air sensor
2
Fault Clearance
2 - 5