Braun Infusomat Space Service Manual page 104

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

3
Disassembly / Assembly
Visual Inspection
Electrical Safety
Functional Inspection
According to IEC 60601-1 or IEC 62353
Air inline sensor
(see TSC for limit values)
Infusomat® Space Line, standard type
❒ Water value
❒ Difference of temperature sensors
❒ Air alarm
❒ Air value
Note: If the water or air value is not
reached immediately, these values have to
be read again after 120 seconds.
Switch on unit without power supply
❒ Self-test
❒ Magnetic function of the battery com-
partment cover
❒ Charging level of the battery module
Remove battery module
❒ Alarm for at least 3 minutes.
(Part 3 of 3)
Date:
Name:
Signature:
3 - 58
Infusomat® Space 6.0