Braun Infusomat Space Service Manual page 122

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

6
Procedural Instructions on the TSC
Functional Inspection Power Supply SP
6 - 10
9. Generate an air bubble of approx. 0.4 ml (approx. 56 mm
length of line with air) in the supply line to the pump. An
alarm is triggered when the air bubble is detected.
10. Insert an air-filled Infusomat® Space Line into the device and
close the operating unit, or disconnect the line at the con-
tainer and "empty", so that there is no longer any water in the
line.
Line change
11. In the
menu, press the ">" button on the device.
The value of the air inline sensor is displayed on the LC dis-
play.
12. Read off value after approximately 10 seconds and compare
with the specifications of the TSC.
13. To exit the Service Info mode press the ">" button once more.
Note
If the water or air value is not reached immediately, these values
must be read again after 120 seconds.
Battery Check
1. Switch device off.
2. Disconnect the power supply.
3. Switch on unit.
Self-test is carried out.
4. Open battery compartment cover during operation.
An alarm is output.
Mechanical Inspection
1. Connect the power supply to a Space system device and
check the lock for proper function.
Functional Check
1. The connected device operates correctly after being switched
on.
Infusomat® Space 6.0