6 - 10
Procedural Instructions on the TSC
6 -
Visual Inspection
EN
Infusomat® Space 6.0
Infusomat® Space
1. Check the Infusomat® Space and accessories for cleanness.
2. Check the Infusomat® Space and accessories for complete-
ness and check configuration.
3. Check the Infusomat® Space and its accessories for damage
and the labels for readability. Pay special attention to the fol-
lowing parts:
a) Safety clamp
b) Pressure sensor membrane, downstream
(not damaged)
c) Pressure sensor membrane, upstream
(not damaged)
d) Pressure webs for pressure sensors
(smooth running, movable without jamming)
e) Pump membrane
(not damaged)
f)
Coding of pump slide guide present
g) Pump connecting rod
(12 pieces, not damaged)
h) Pump slide
(12 pieces, smooth running and engaged)
i)
Air inline sensor
(free of oil and ultrasound gel),
free of cracks)
j)
Pressure webs for air inline sensor
(smooth running, movable without jamming)
k) Drop sensor
(lens)
l)
Screw covers
m) Connectors "P2" and "P3"
n) Completeness and correct position of the insulating
washers on the operating unit back panel
o) Completeness and readability of both pictograms on the
inside of the operating unit