4
Servicing the Unit
Fig.: 4 - 2
Connector corroded
Fig.: 4 - 3 Example: connector on SpaceStation
4 - 2
Connector cleaned
Internal tube routing areas
The slide guide and the internal tube guides can be cleaned with
disinfectants containing 1-propanol or didecyl dimenthyl ammo-
nium chloride.
The slide guide can be prised out with a pointed object (ballpoint
pen) through a notch in the housing front panel on the right lower
side
(Fig.: 4 -
2). The cover can then be cleaned by immersing it,
the pump slide can be cleaned with a soft cloth.
When you reinsert the slide guide, make sure it is not damaged
and audibly snaps into place.
Cleaning of the connector
Corroded or soiled connector contacts can result in connection
problems and fault messages output during operation. After it has
been identified by a visual or function check, clean the offending
connector as follows.
Cleaning agents and materials:
-
Contact cleaner: Kontakt WL, order no. 20.710.6.13.12.01 or
DeoxIT® D5 Spray, part no. D5S-6
-
Contact spray: Kontakt 60 plus, order no. 20.739.6.09.12.01
or DeoxIT® D100 Spray, part no. D100S-2
-
Cleaning brush: Kontakt Chemie, ref. no. 30206 (3 each) or
Connector Cleaning Brush, part no. AB-25 (25 each)
-
Swabs: Chemtronics Chamois Tips CC50 (50 each) or DeoxIT®
Precision Foam Swabs, part no. SWPX-25 (25 each)
*
Not all products are available in all regions.
Caig Laboratories, Inc. (www.caig.com)
*
Infusomat® Space 6.0