Braun Infusomat Space Service Manual page 34

Infusion pump
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

2
Unit Diagnosis / Calibration
Error Code
1
2100
Timebase too fast
2
2101
Timebase too slow
3
2102
Timebase failure
4
2103
Keyboard High
5
2104
ON/OFF key pressed too long
6
2105
No keydecode
7
2106
ROM Romtest defective
8
2107
ROM program defective
9
2108
FuP has not changed active reset value
10
2109
MPU_Test failed
11
2110
RAM_Test failed
12
2111*
KUP active reset
13
2112 2114
Internal Error
14
2115
Drive too fast
15
2116
Drive too slow
16
2117 ... 2118
Internal Error
17
2119
LCD backlight on defective
18
2120
LCD backlight off defective
19
2121
Red LED on defective
20
2122
Red LED off defective
21
2123
Key pressed too long (without ON/OFF
key)
22
2124 ... 2127 Internal Error
23
2128
Drive motion incorrectly forward
24
2129
Drive motion incorrectly backward
25
2130 ... 2159 Internal Error
26
2160 ... 2162 PCA Error
27
2170 ... 2173 DLOCK Error
Table 2 - 3 Error Codes of the Control Microprocessor (Part 1 of 3)
2 - 4
Definition
Quartz of the processor PCB
Quartz of the processor PCB
Quartz of the processor PCB
Keyboard defective
Keyboard defective
Keyboard defective
Software
Pump was not switched off cor-
rectly
Motor drive
Detection of direction of rotation
LC display defective
Keyboard defective
Motor drive
Detection of direction of rotation
Error in relation to PCA therapy
together with special settings
Error in relation to Datalock to-
gether with special settings
Error Codes of the Control Microprocessor
Possible Cause
* Note: If alarm 2111 occurs during infusion, the device history logs may not be re-
trievable.
*
Fault Clearance
Exchange processor PCB
Exchange processor PCB
Exchange processor PCB
Exchange operating unit
Exchange operating unit
Exchange operating unit
Update unit software
Restart device
Exchange processor PCB
Exchange operating unit
Exchange operating unit
Replace processor PCB
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Infusomat® Space 6.0